| Technical Specifications |
|
Test Voltage 500 V DC |
Range: 200 kW to 10 GW
Accuracy: 5%
Range: 10 GW to 100 GW
Accuracy: 20% |
|
Test Voltage 1000 V DC |
Range: 200 kW to 20 GW
Accuracy: 5%
Range: 20 GW to 200 GW
Accuracy: 20 |
|
Test Voltage 2500 V DC |
Range: 200 kW to 50 GW
Accuracy: 5%
Range: 50 GW to 500 GW
Accuracy: 20% |
|
Test Voltage 5000 V DC |
Range: 200 kW to 100 GW
Accuracy: 5%
Range: 100 GW to 1 TW
Accuracy: 20% |
|
Short circuit current |
> 1 mA
and < 2 mA |
|
Bar graph |
Range: 0 to 1 TW |
|
Insulation test voltage |
Accuracy: 0% to 10% at 1 mA load current |
|
Induced AC mains current rejection |
Max: 2 mA |
|
Charging rate for capacitive load |
5 s/µF |
|
Leakage current |
Range: 1 nA to 2 mA
Accuracy: ±(5% + 2 nA) |
|
Capacitance measurements |
Range: 0.01 µF to 15.00 µF
Accuracy: ±(15% rdg + 0.03 µF) |
|
Live circuit indicator |
Range: 30 V to 600 V (AC/DC, 50/60 Hz)
Accuracy: ±(5% + 2V) |
|
Timer |
0 - 99
minutes
Settable: 1 minute increments
Indication: 1 second resolution |
|
Ramp |
0% to
100% of selected test voltage, or until breakdown |
|
| |
| |
| Environmental Specifications |
|
Operating Temperature |
-20°C to
+40°C |
|
Storage Temperature |
-20°C to
+65°C |
|
Humidity (Without Condensation) |
80% at
31°C decreasing linearly to 50% at 40°C |
|
Dust/water resistance |
IP-40 |
|
Operating Altitude |
2,000 m |
|
| |
| |
| Safety Specifications |
|
Safety Rating |
EN
61010-1, CAT III 600V, EN61557, Parts 1 & 2 |
|
EMI immunity |
EN 61326 |
|
| |
| |
| General Specifications |
|
Size |
170 x
242 x 333 mm |
|
Weight |
4 kg |
|
Batteries |
12 V,
lead-acid, rechargeable |
|
Warranty |
2 years |
|
Battery life |
Test voltage Number of tests
500 V 3600
1000 V 3200
2500 V 2500
5000 V 1500 |
|
Shock & Vibration |
MIL-PRF-28800F, Class II |
|
Charger input |
AC: 85 to 250 V AC, 50/60 Hz, 50 VA
DC: 10 to 15 V DC, 30 VA ( (car adapter) |
|
| |
| |
| Rental Inquiry Information |
| |
|
Reference: |
1550B |
|
Description: |
Fluke 1550B 5kV Insulation
Tester |
|
| |