Temperature Test and Measurement
The DPI820 Thermometer has two inputs T1 and T2 and allow individual
measurement or simultaneous comparisons. For differential
applications; e.g. heating boiler bypass, T1 minus T2 can be
displayed.Interchangeable Probes and Wide Thermocouple
compatibility allow general temperature measurement or the use
of special sensors for demanding applications; e.g. high
temperature, hygienic and aggressive media.
Data Logging Memory: 1000 single or 750 dual screens with
date and time, with six Header Tags for user characters to identify
groups of readings. The RS232 port is 19.2k baud, 8-data bits, 1
stop bit, no parity, Xon/Xoff.
|

|
Features
- Single, dual or differential input readings.
- Wide probe compatibility for general measurement and
specialist applications.
- Advanced Cold Junction compensation virtually eliminates
errors due to ambient temperature changes.
- Data logging with adjustable period or manual recording.
- Large backlit display, menu driven interface.
- RS232 interface for PC or serial printer.
- Robust and weatherproof.
- Convenient, one hand operation.
- Secure grip, impact resistant, elastomer protected.
Applications
- Temperature test and measurement.
- Batch sampling and environmental monitoring.
- Comparison test against reference sensor.
- Laboratory standard.
Druck DPI 820 Specifications (PDF)
(documents are in PDF format and will open in a new window) |
|
| Functions |
| Type |
°C (°F) |
| Thermometer |
Dual input T1, T2 and T1-T2 |
| Pressure Module |
IDOS Universal Measurement Module (optional refer to DOS
datasheet) |
| Instrument Functions |
Hold, scaling, max/min/avg, filter, alarm, tare |
| Flow Scaling |
25 pressure units, flow scaling, leak test (when fitted with
IDOS pressure module) |
| Memory |
1000 point data memory and RS232 |
|
| |
IDOS Flexibility (Intelligent Digital Output Sensor)
Universal pressure modules are available from 10 in H2O to
10,000 psi (25 mbar to 700 bar)Total Flexibility
IDOS modules can be used with any compatible instrument; for example, a
DPI 820 thermometer can become a fully featured pressure indicator or can
read pressure and temperature simultaneously.
Plug and Play
Modules are interchangeable between any of the DPI 800 Series instruments,
requiring no set-up or instrument calibration.
Please refer to IDOS UPM data sheet. |
| |
| DPI 820 Specifications |
| *Accuracy includes operation from 10°C
to 30°C (50°F to
86°F), one year stability and calibration
uncertainty. |
|
Type |
Standard |
*Accuracy |
Range |
|
K |
IEC 584 |
0.6°C
(11°F) |
-170°C to
1370°C (-454°F
to 2498°F) |
|
J |
IEC 584 |
0.5°C
(0.9°F) |
-210°C to
1200°C (-346°F
to 2192°F) |
|
T |
IEC 584 |
0.3°C
(0.6°F) |
-270°C to
400°C (-454°F
to 752°F) |
|
B |
IEC 584 |
1.0°C
(1.8°F) |
50°C to 1820°C
(122°F to 3308°F) |
|
R |
IEC 584 |
1.0°C
(1.8°F) |
-50°C to
1769°C (-58°F
to 3216°F) |
|
S |
IEC 584 |
1.4°C
(2.5°F) |
-50°C to
1769°C (-58°F
to 3216°F) |
|
E |
IEC 584 |
0.4°C
(0.7°F) |
-270°C to
1000°C (-454°F
to 1832°F) |
|
N |
IEC 584 |
0.6°C
(1.1°F) |
-270°C to
1300°C (-454°F
to 2372°F) |
|
L |
IEC 43710 |
0.3°C
(0.6°F) |
-200°C to
900°C (-328°F
to 1652°F) |
|
U |
IEC 43710 |
0.3°C
(0.6°F) |
-200°C to
600°C (-328°F
to 1112°F) |
|
C |
|
1.0°C
(1.8°F) |
0°C to 2320°C
(32°F to 4208°F) |
|
D |
|
1.0°C
(1.8°F) |
0°C to 2495°C
(32°F to 4523°F) |
|
| Accuracies quoted for mid range and do not include sensor error |
| |
| |
| Rental Inquiry Information |
| |
|
Reference: |
DPI820 |
|
Description: |
Druck DPI820 Dual Input
Thermocouple Thermometer |
| |
|
|